Suitable to measure temperature, resistance and voltage.
Setting of chracteristic values with a software for visualization via HART protocol.
- Input: resistance thermometer / thermocouple /
voltage (mV)/ resistance (potentiometer)
- Output: 4...200 mA current loop, HART protocol
- Supply: 15...45 VDC out of current loop
- Accuracy: 0,08...0,12% of range, 0,2...2 K
0,1...7,5 ohms, 10...50 µV
- Indication: LCD display (backlighting)
- Galvanical insulation: input / output
- Settings: with software via HART-protocol
- Enclosure: diecast aluminium
|
|
Programmable to customize to the installation
- Input: TC / RTD / current / voltage /
resistance / potentiometer
- Output: current / voltage / relay points
- Voltage supply: 21,6 ...253 VAC
19,2...300 VDC
- Mounting: on DIN - top hat rail / wall
- Supply for 2-wire transmitter
galvanically insulated
- Accuracy: 0,1% of span
- LED indicator: normal operation / malfunction / relay active
- Programming: with programming front
|
|
in top hat rail-case, for sensors, universal
programmable, for general applications
- Input resistance thermometer -100...+850°C,
thermocouple -100...+1800°C, resistance and potentiometer 10ohm...10kOhm
- PC - configurable with "DTV - Control"
programming software
and interface
- Output 0...20mA, -10...+10V
pulse or relay output
- Supply voltage 24, 115, 230, 400VAC
or 12...50VDC
- Galvanic separation
|
|
for mounting DIN terminal head form B
Output 4...20 mA, programmable
- Input: thermoelement or
resistance thermometer
- Output: 4...20 mA current loop
- freely programmable features
- EEx version ia IIC T4...T6
- with setup program for windows
- Accuracy up to +/- 0,2 K
- Degree of protection IP 54/66,
when using terminal head
- for use with terminal head form B
|
|
for mounting in DIN terminal head form B
Output 4...20 mA or 0...10 V, programmable
- Resistance thermometer
- Output: 4...20 mA current loop / 0...10 V
- Supply with output voltage: 15...30 VDC
- freely programmable features
- Measuring circuit monitoring
- With setup program for windows
- Accuracy 0,2 % FS / 0,2 K
- For use with terminal head form B
|
|
head mounting, configurable ranges selectable by solder bridges
for Pt100 resistance thermometers and thermocouples
- Input for Pt100 2- or 3-wire or
thermocouples type T, J, K or S
- Configurable ranges free selectable
by solder bridges
- Analog output 4...20mA,
2-wire design
- Fault signal for sensor burnout
and sensor short circuiting
- Mounting in connecting head DIN B
or with adaptor on top hat rails
- Optionally: EX-protection,
II 1G EEx ia IIB / IIC T4 / T5 /T6
|
|
head mounting, universally programmable for resistance thermometers,
thermocouples, resistance sensors and mV-sensors
- Input: resistance thermometer
2- 3- or 4-wire
thermocouples type T, E, J, L, K, N, U, R, S, B, W3 or W5
mV-sensors up to 800mV
- Configurable for PC, MS-Windows from 3.11
operating surface "DTM-Control"
- Analog output 4...20mA, reversible,
2-wire technics
- Temperature linear output for input signal
for resistance thermometers and thermocouples
- Galvanic insulation
- Optional EX-protection, intrinsically
safe EEx ia IIC T4 / T5 / T6 or
EEx ib IIC T4 / T5 / T6
|
|
for head mounting, for use as robust
limit value switch or controller
- Input for resistance thermometers
or thermocouples
- For connecting heads B-HD / BUS / BUZ
- Operating with all temperature sensors
as "Stand alone" device
- Continuously adjustable limit value
with variable hysteresis (250VAC/5A)
- Indication status of relay (LED red)
- Accuracy below +/- 0,5% of end of scale value
|
|